Code: SC101
Description: Exposure lamp failure - Timing & Condition - • The lamp output check signal (LAMPDET) is still HIGH 250 ms after the lamp turns on (the ON check is canceled if the lamp goes off within 250 ms). • The lamp output check signal is still LOW 30ms after the lamp turns off (the OFF check is canceled if the lamp turns on within 30 ms).
Causes: • Blown lamp • Blown thermostat • Blown fuse (FU301) • Defective lamp regulator • Poor connection • Defective harnesses • Defective AC drive board • Defective sub/main scanner IPU board
Remedy: 1. Visually check the lamp element or check both ends of the lamp terminals with a multi-meter. 2. Check the current at the ends of the thermostat terminals with a multi-meter. 3. Check if the connectors (CN1, CN2, and CN3) on the lamp regulator are properly connected. 4. Check the current through the 3 harnesses. 5. If 100Vac is provided from the AC drive board (check CN3-1 and 5 on the lamp regulator) replace the lamp regulator. 6. If 100Vac is not supplied at CN3, Replace the AC drive board. 7. Replace the sub scanner IPU board and/or main scanner IPU board [Signal Check] • LAMPDET: CN2-2 on the lamp regulator/ CN403-A10 on the sub scanner IPU board • LAMPTRIG: CN403-A9 on the sub scanner IPU board/CN2-3 on the lamp regulator