Code: 010032
Description: Fan Test Failed -Current not detected in one or bothfans. Error data (XX):
1 -Electronics Module fan not running. 2 -Service station fan not running. 3 -Bothfans not running. 10 -Electronics Module fan is short-circuited. 20 -Service Station fan is short-circuited. 30 -Bothfans are short-circuited. 12 -Service Station fan is not running and the Electronics Module fan is short-circuited. 21 -Electronics Module fan is not running and the Service Station fan is short-circuited.
1 -Electronics Module fan not running. 2 -Service station fan not running. 3 -Bothfans not running. 10 -Electronics Module fan is short-circuited. 20 -Service Station fan is short-circuited. 30 -Bothfans are short-circuited. 12 -Service Station fan is not running and the Electronics Module fan is short-circuited. 21 -Electronics Module fan is not running and the Service Station fan is short-circuited.