Description: APS pressurize error 01. Run IDS diagnostic. If problem persists, call your support representative. The air system is unable to pressurize to the expected working pressure, which, depending on the actual working cycle, could be 5.2–5.6 psi (normal working pressure) or 3.6–4 psi (during refilling the non-refilled side is pressurized to prevent backwards refill).
Causes: a. Air tube disconnected from the air system b. Air pump failure (pump or ISS PCA failure) c. Air pressure bottle leak d. Air tube leak (tube damaged) e. Air relief valve failure f. Intermediate tank leak g. Air sensor failure, leading to ISS main board failure h. Intermediate tank supports failure
Remedy: 1. Review all air tube connections and ensure that there is no disconnection (tubes from pumps and relief valve to tap, tubes from tap to intermediate tanks). 2. Perform the front and rear air pumps diagnostic procedures. See Table 5-10 Ink delivery system. 3. If the pumps do not work, check pump connections and verify pin integrity. 4. If connections are correct, connecting to another working port and start the pump. 5. If the pump works with another port, replace the ink system main board. 6. If the pump fails, replace the air pump. See ISM pump and relief valve. 7. Check that the pressure bottle cap is tight and there is no leakage. 8. Pressurize the system and check for any failed tube. Leave the pumps on test with diagnostic procedure , the pump will operate until 6 psi is reached. 9. Check relief valve is closed and air is not escaping from it when pressurizing. Swap the relief valve from the other side to verify that it is the relief valve that is failing. If the relief valve fails, replace it. See ISM pump and relief valve. 10. Swap all intermediate tanks and ensure that there is no leak. If a leak is detected, swap the intermediate tanks again one by one until the leak is discovered. Replace the intermediate tank as appropriate and force-fill intermediate tanks. See ISM intermediate tank. 11. Swap air pressure sensor tube connections to the PCA and display values if required. If the sensors do not have similar behavior, replace the ISS PCA. See ISS PCA. 12. Try to adjust tubes and check connections, clean if required. 13. If necessary, replace the ink circuit module. See ISM ink circuit module.