Code: E2-10
Description: System Console Drive Motor Rotation
Causes: 1. Drive Mechanism defective. 2. Drive Motor connector disconnected. 3. Drive Motor defective. 4. CST2 PCB connector disconnected. 5. CST3 PCB connector disconnected. 6. CST3 PCB defective. 7. PS connector disconnected. 8. PS defective. 9. SPC PCB connector disconnected. 10. SPC PCB defective.