Code: 6790
Description: AFE offset adjustment error
During the offset adjustment, the offset value does not fall within the predetermined range a total of three times including retries.
During the offset adjustment, the offset value does not fall within the predetermined range a total of three times including retries.
Causes: • Scanner unit • Controller board (CTLB)
Remedy: 1 Check the connector between scanner unit- CTLB J101 for proper connection and correct as necessary. 2 Check for possible extraneous light and correct as necessary. 3 Clean the lens, mirrors, CCD surface, and shading sheet if dirty. 4 Correct reflective mirror of the scanner if faulty, or change scanner mirror. 5 Replace scanner unit 6 Replace CTLB