Code: 9401
Description: Exposure lamp lighting failure detected
The intensity of the lamp is not stabilized within a predetermined period of time during the lamp stabilization check process in a lamp warm-up cycle.
The intensity of the lamp is not stabilized within a predetermined period of time during the lamp stabilization check process in a lamp warm-up cycle.
Causes: • Controller board (CTLB) • Scanner unit
Remedy: 1 Check the connector between scanner unit-CTLB J101 for proper connection and correct as necessary. 2 Check for possible extraneous light and correct as necessary. 3 Clean the lens, mirrors, CCD surface, and shading sheet if dirty. 4 Correct reflective mirror of the scanner if faulty, or change scanner mirror. 5 Replace scanner unit 6 Replace CTLB