Code: C8302
Description: Cooling fan trouble
<When DF-624 is installed> <When DF-701 is installed> • The lock signal continues to detect H during a given time while the DF cooling fan motor (FM1) is spinning. • The lock signal continues to detect L during a given time while the cooling fan (FM1) is during halts.
<When DF-624 is installed> <When DF-701 is installed> • The lock signal continues to detect H during a given time while the DF cooling fan motor (FM1) is spinning. • The lock signal continues to detect L during a given time while the cooling fan (FM1) is during halts.
Remedy: When DF-624 is installed: 1 Check the connector between FM1-DFCB J16 for proper connection and correct as necessary. 2 Check the fan for possible overload and correct as necessary. 3 FM1 operation check DFCB J16-3 (ON) DF-624 3-B 4 Replace FM1. 5 DFCB F6 conduction check. 6 Replace DFCB.
When DF-701 is installed: 1 Check the connector between FM1-DFCB J16 for proper connection and correct as necessary. 2 Check the fan for possible overload and correct as necessary. 3 FM1 operation check DFCB J16-3 (ON) DF-701 7-G. 4 Replace FM1. 5 DFCB F6 conduction check. 6 Replace DFCB.
When DF-701 is installed: 1 Check the connector between FM1-DFCB J16 for proper connection and correct as necessary. 2 Check the fan for possible overload and correct as necessary. 3 FM1 operation check DFCB J16-3 (ON) DF-701 7-G. 4 Replace FM1. 5 DFCB F6 conduction check. 6 Replace DFCB.