Code: C0954
Description: 4th Drawer Lift-Up Motor malfunction
Remedy: 1 4th Drawer Lift-Up Motor turns when the 4th Drawer is slid out and then back in NO Correct drive coupling. 2 4th Drawer Lift-Up Motor operation: the voltage across PJ10A-1 on the Master Board and GND is DC0V (motor deenergized) and DC24V (motor energized) after the 4th Drawer has been slid in position. YES Change motor. NO Change Master Board. 3 4th Drawer Lift-Up Motor Pulse Sensor operation: the voltage across PJ9A-9 on the Master Board and GND varies in the range between DC0V and DC5V while the 4th Drawer Lift-Up Motor is energized. YES Change Master Board. NO Change pulse disk and gear. Change sensor. 4 4th Drawer Lift-Up Sensor operation: the voltage across PJ9A-2 on the Master Board and GND is DC5V (sensor unblocked) and DC0V (sensor blocked). YES Change Master Board. NO Correct retraction mechanism. Change sensor.