Code: 0001
Description: Transport motor malfunction
• The motor lock signal remains HIGH for a predetermined consecutive period of time while the transport motor remains energized. • The rack positioning sensor is in the blocked state when the rack motor remains deenergized. • The rack positioning sensor is not blocked a second time even after the lapse of a predetermined period of time after it has been blocked once while the rack motor remains energized.
• The motor lock signal remains HIGH for a predetermined consecutive period of time while the transport motor remains energized. • The rack positioning sensor is in the blocked state when the rack motor remains deenergized. • The rack positioning sensor is not blocked a second time even after the lapse of a predetermined period of time after it has been blocked once while the rack motor remains energized.
Remedy: 1 Check the M1 connector for proper connection and correct as necessary. 2 Check M1 for proper drive coupling and correct as necessary. 3 Check the PRCB connector for proper connection and correct as necessary. 4 M1 operation check. PRCB PJ8-1 to 6 1600W: C-2 1650EN: C-2. 5 Change PRCB. 6 Change DCPU.