Code: F11
Description: Communications error between system-CPU and scanner-CPU
Remedy: Did the error occur immediately after updating the firmware? Y: Are the proper files saved in the PC? (In case of updating by PC) Are the type, attaching position and direction of the ROM correct? (In case of updating by download jig) N: After correcting, perform the firmware updating again. N: 1. Check if the IMG board connector J161 is disconnected. 2. Check if the SCM board connector J1 is disconnected. 3. Check if the harness between the IMG and SCM boards is open-circuited. 4. Check if the IMG board is securely connected to the MTH1 board. 5. Check if the SYS board is securely connected to the MTH1 board. 6. Check the version of FROM on the SYS board. 7. Check the version of FROM on the SCM board. 8. Check the scanner related setting values (08-360 and 361). 9. Replace the SYS board. 10. Replace the IMG board. 11. Replace the SCM board.