Code: C260
Description: Peak detection error: Lighting of the exposure lamp (white reference) is not detected when power is turned ON, or communication error with CCD board.
Causes: Lens unit SYS board Exposure lamp Power supply harness
Remedy: 1 Does the exposure lamp light? (Perform the output check: 03-267) Yes It is lit. 2 No It is not lit. 3 2 Shading correction plate Check if there is any scratch or stain on the shading correction plate. Mirror Check if the mirror is tilted. • Check that the lens is reflected in the mirror looking at carriage-1 from the upper position. • Check that the mirror is secured at the leaf spring. Exposure lamp 1. Check if the exposure lamp is correctly lit. 2. Check if the harness is connected properly to the exposure lamp connector. 3. When the carriage is driven, check if the harness interferes with it or parts are caught in it. CCD board / Lens unit 1. Check if the connector of the CCD board is connected properly. 2. Check if the CCD board is installed properly. (Check that the lens unit is not tilted or the screw is securely tighten.) SYS board 1. Check if the connector of the SYS board (CN122, CN126) is connected properly. 2. Check if the mounted parts on the SYS board are damaged or abnormal. 3. Check if 10 V is output from the power supply for CCD. (CN122-1pin). 3 SYS board 1. Check if the supply cable is connected properly to the connector (CN127). 2. Check if the mounted parts on the SYS board are damaged or abnormal. Exposure lamp 1. Check if the harness of the exposure lamp is connected to the LED light source properly. 2. Check if the exposure lamp is scratched or damaged. 3. Check if the exposure lamp harness comes off the board. Power supply harness 1. Check if wiring of the power supply harness (CN127) is abnormal. 2. Check if the harness is scratched or open circuited.