Description: Scanner fuse blowout: 24V power for the scanning system is not supplied at the scanner warming-up after power-ON.
Causes: SYS board Fuse (F204) Power supply harness LVPS
Remedy: Is 24V supplied to the SYS board? Yes Check the following because the signal for checking 24V on the SYS board is abnormal. 1. Check if 3V is input in 35 Pin of the scanner CPU (IC301). 2. Check if the mounted parts on the SYS board are damaged or abnormal. 3. Check if 24V on the SYS board is short circuited. 4. Check if 24V is supplied to the SYS board (CN127). No 1. Check if the 24V supply harness is properly connected to the connector (CN127). 2. Check if 24V and SG on the SYS board are short circuited. 3. Check if the power supply is short circuited by pulling out the supply harness on the SYS board (CN127). 4. Check if the fuse on the LVPS (F204) is open circuited. 5. Check if there is no abnormality on the LVPS.