Code: C900
Description: Connection error between SYS board and LGC board
Remedy: (1) Check if the connector CN423, CN424 on the IMG board and the connector CN319, CN320 on the LGC board are completely inserted. (2) Check if the connector pin between the IMG board (connector CN423, CN424) and the LGC board (connector CN319, CN320) is disconnected. (3) Check if the connector CN422 on the IMG board and the connector CN135 on the SYS board are completely inserted. (4) Check if the connector pin between the IMG board (connector CN422) and the SYS board (connector CN135) is disconnected. (5) Check if the conductor patterns on the IMG board, LGC board and SYS board are short circuited or open circuited. (6) Replace the LGC board if no problem is found in steps from (1) to (5) above. (7) If the problem is not corrected with the replacement of the LGC board, reinstall the removed LGC board and replace the IMG board. (8) If the problem is still not corrected with the replacement of the IMG board, reinstall the removed IMG board and replace the SYS board.