Code: EE EU
Description: Toner concentration reference control level setup trouble (Undertoner)
Causes: "When setting the toner concentration reference control level , the toner concentration sensor output is not within the specified range. (The output level is179 or above.)"
Remedy: Case 1 Trouble position/cause Toner density sensor trouble / Toner density sensor signal line connection trouble Remedy (Check) Check the toner concentration output level with SIM 25-1. (Repair) Toner concentration sensor replacement (Developing unit replacement) / Power Source-ON (After-work) Execute ADJ M3. Note Case 2 Trouble position/cause Process control PWB trouble Remedy (Check) Check the toner concentration sensor output level with SIM 25-1. (Repair) Process control PWB replacement / Power Source-ON (After-work) Note Case 3 Trouble position/cause PCU MAIN PWB trouble Remedy (Check) Check the toner concentration sensor output level with SIM 25-1. (Repair) PCU MAIN PWB replacement / Power Source-ON (After-work) Re-enter the setup values and adjustment values. (Install the EEPROM of the defective PCU MAIN PWB to a new PCU MAIN PWB.)