Causes: Abnormal value is read from the RTC on the ICU PWB.
Remedy: Case 1 Cause RTC circuit abnormality Check & Remedy Set the time again with the key operation, and check that time advances properly. Check the RTC circuit. Case 2 Cause Battery voltage fall. Battery circuit abnormality. Check & Remedy Check to insure that the battery voltage is about 2.5V or above. Check the battery circuit.