Description: Adjustment value check sum error (SCU detection) EEPROM (SCU) check sum error
Causes: Control circuit hang-up due to electrical noises EEPROM trouble SCU PWB EEPROM access circuit trouble
Remedy: Case 1 Cause Control circuit hang-up due to electrical noises Check and remedy Use SIM16 to cancel U2 trouble. Case 2 Cause EEPROM trouble Check and remedy To avoid deletion of the counter data/adjustment data, use the simulation to save the counter/ adjustment values. (If there is a printer option, execute SIM22-1 to save the counter data and the adjustment values.) Replace the SCU PWB. Use SIM16 to cancel the trouble. Case 3 Cause SCU PWB EEPROM access circuit trouble Check and remedy Check that EEPROM is properly set.