Remedy: 1. Check the installation of the CTD Sensor Assembly. Is the CTD Sensor Assembly installed properly? Go to step 2. Reinstall the CTD Sensor Assembly. 2. Check the connection between the CTD Sensor Assembly and MCU PWB. Is the connector P/J4561 connected securely? Go to step 3. Connect the connector P/J4561. 3. Check the conductivity between the CTD Sensor Assembly and MCU PWB. Is the connection between P/J4561 <=> P/J456 conducting properly? Go to step 4. Replace the Main CTD Harness Assembly. 4. Check the connection in the CTD Sensor Assembly. Is the connector P/J4563 connected securely? Go to step 5. Connect the connector P/J4563. 5. Replace the CTD Sensor Assembly. Does the error persist? Replace the MCU PWB. Troubleshooting complete.