Description: Program Faults A software error has occurred. The following troubleshooting procedure applies to these errors: 7,0xx.6x The system detected a program fault during operation.
Causes: NVRAM, PL 5.0
Remedy: • Reset NVRAM and retest. • If the problem persists, follow the procedure below. 1 Reseat all Electronics Module connections. Does the problem persist? Go to Step 2. Complete. 2 Check ground integrity. Are the system grounds connected? Reset NVRAM. If the error persists, replace the NVRAM device. Reconnect the system grounds.