Description: Scanner Calibration Faults RAP 362-450-00 Dark range status bit was not clear prior to calibration. 362-451-00 Dark range status bit was not set after calibration. 362-452-00 Pixel offset status bit was not clear prior to calibration. 362-453-00 Pixel offset status bit was not set after calibration. 362-454-00 Gain range status bit was not clear prior to calibration. 362-455-00 Gain range status bit was not set after calibration. 362-456-00 Pixel gain status bit was not clear prior to calibration. 362-457-00 Pixel gain status bit was not set after calibration. 362-458-00 Highest intensity image pixel value exceeded maximum tolerance. 362-459-00 Pixel offset error (High) exceeded maximum adjustment allowed during dark calibration. 362-460-00 Pixel offset error (Low) exceeded maximum adjustment allowed during dark calibration. 362-461-00 Highest intensity image pixel value was lower than the minimum tolerance. 362-462-00 Pixel gain error (High) exceeded maximum adjustment allowed during white calibration. 362-463-00 Pixel gain error (Low) exceeded maximum adjustment allowed during white calibration. 362-464-00 The scanner was requested to perform another operation while busy. 362-466-00 Pixels were out of range during black calibration. 362-467-00 Pixels were out of range during white calibration. 362-468-00 Pixel clock error from the full width array. 362-469-00 Calibration ASIC comm’s error. 362-470-00 Unable to read the registers in the calibration ASIC. 362-471-00 Calibration ASIC write buffer was full. 362-472-00 Calibration ASIC comm’s time-out.
Remedy: Perform ADJ 60.3 IIT Registration, Magnification and Calibration. The fault persists. Y↓N→Perform SCP 5 Final Actions. Go to Flag 1. Disconnect, then check the scan carriage data ribbon cable and connectors, P/ J445 on the Scanner CCD PWB and P/J412 on the Scanner PWB. The ribbon cable and connectors are clean and undamaged. Y↓N→Clean or repair the scan carriage data ribbon cable or connectors. Install new components as necessary: • Scan carriage data ribbon cable, PL 60.20 Item 17. • Scan carriage assembly, PL 60.20 Item 1. • Scanner PWB, PL 60.20 Item 4. Reconnect the scan carriage data ribbon cable. Go to Flag 2. Disconnect, then check the SBC PWB/scanner PWB data cable and connectors, P/J411 on the Scanner PWB and P/J854 on the SBC PWB. The data cable and connectors are clean and undamaged. Y↓N→Clean or repair the ribbon cable or connectors. Install new components as necessary: • SBC PWB/scanner PWB data cable, PL 60.20 Item 22. • Scanner PWB, PL 60.20 Item 4. If the fault persists, perform the 303D SBC PWB Diagnostics RAP. Perform dC301 Scanner System NVM Initialization. Then perform ADJ 60.3 IIT Registration, Magnification and Calibration. If the fault persists, perform an AltBoot, GP 4.