Code: 620
Description: Communication error between BICU and ARDF
The BICU does not receive a response from the ARDF main board for 4 seconds or more. The BICU receives a break signal from the ARDF main board.
The BICU does not receive a response from the ARDF main board for 4 seconds or more. The BICU receives a break signal from the ARDF main board.
Causes: • Poor connection between the BICU and ARDF main board (DF connector) • ARDF main board defective • BICU defective