Code: 601
Description: Communication error between BICU and scanner unit
Within 800 ms after power on, after 3 attempts the BICU does not communicate with the SIB via the serial line.
Within 800 ms after power on, after 3 attempts the BICU does not communicate with the SIB via the serial line.
Causes: • Serial line connecting the BICU and SIB defective • External noise on the serial line • SIB board defective • BICU board defective