Code: E614 0601
Description: Flash PCB error. An error was detected in the license-related area. (Initialization failed at startup or I/O error at startup) When this error occurs, the system has not been started normally. Therefore, it may not be recorded in the error log.
Remedy: [Related parts] • Flash PCB (UN61) • Main Controller PCB (UN05) (Unit of replacement: MAIN CONTROLLER PCB ASS'Y) [Remedy] Perform the following in the order while checking whether the error is cleared. • Reinstall the necessary application software and restore the backup data once the error is cleared. 1. Obtain the necessary backup data by referring to the backup data list. 2. Enter safe mode using (2+8) startup, and execute [4] Clear/ Format> [2] Flash Format (Flash format) using a USB memory. 3. After replacing the Flash PCB, reinstall the system software using SST or a USB memory. 4. Replace the Main Controller PCB. [Reference] For backup and restoration, refer to "Adjustment> Actions after Replacement" and "Appendix> Backup Data List" in the Service Manual.