Code: E211
Description: The scanning lamp (fluorescent lamp) thermistor has an open circuit.
• At a 0°C or lower condition, the temperature fails to reach 10°C after supplying the scanning lamp heater with power for 2 min.
• At a 0°C or lower condition, the temperature fails to reach 10°C after supplying the scanning lamp heater with power for 2 min.
Remedy: 1) Disconnect connector J115 of the reader controller PCB, and measure the resistance between connectors J115-5 (GND) and J115-6 (FL-THE) on the harness side. Is it 10 W or less or 100k W or more? NO: The reader controller PCB is faulty. Go to step 4). YES: Go to step 2). Lamp heater 2) Disconnect connector J2039 of the lamp heater, and measure the resistance between connectors J2039-2 (GND) and J2039-3 (FLTHE) of the lamp heater. Is the resistance 100 W or less or 100k W or more? YES: Replace the lamp heater. Flat cable and connector (poor contact) 3) Are connectors J115 of the reader controller PCB, connectors J2037 and J2035 of the flat cable, and connector J2039 of the lamp heater connected securely? NO: Connect them securely. Reader controller PCB 4) When the machine is turned on, is the voltage between HJ122-1 (GNDU) and J122-2 (+38 V) on the harness side of the connector of the reader controller PCB 38 V? NO: See “The reader unit DC power supply is absent.” YES: Replace the reader controller PCB.