Code: E220
Description: The scanning lamp (fluorescent) is faulty.
• The light of the scanning lamp fails to reach a specific level of intensity within 10 sec after it has been turned on.
• The light of the scanning lamp fails to reach a specific level of intensity within 10 sec after it has been turned on.
Remedy: Scanning lamp 1) Is the scanning lamp mounted correctly? NO: Mount the scanning lamp correctly. (Caution) 2) Try replacing the scanning lamp. Is the problem corrected? YES: Replace the scanning lamp. Inverter PCB 3) Are connectors J1002 and J1003 on the inverter PCB and connector J129 on the reader controller PCB connected securely? NO: Connect the connectors securely. YES: Replace the inverter PCB. Caution: • If you have to remove the scanning lamp and mount it back, be sure to execute the following in service mode: Function>MISC-R>USELAMP. • If you have mounted a new scanning lamp, be sure to execute the following in service mode: Function>MISC-R>LAMP-ADJ for light intensity adjustment and then Function>CCD>CCD-ADJ.