Description: The rise in temperature of the scanning lamp heater is faulty.
During the initial activation after power-on, the reading does not reach 70 deg C within 5 min after the heater is turned on. Or, during standby or reading operation, the reading does not reach 75 deg C within 3 min after the heater is tuned on.
Remedy: - Check if the connecter (J852, J853) of the light control PCB is mounted properly. - Replace the scanning lamp heater. - Replace the light control PCB. - Replace the reader controller PCB.