Code: 841.00...842.00
Description: 841.00 Image pipeline ASIC 841.01 Service image processing unit: AFE register mismatch 841.03 Service image processing unit: SCNCHN DMA 841.04 Service image processing unit: new line too soon interrupt 841.07 Service image processing unit: tonal data overrun 841.09 Service image processing unit: image processing too slow 841.96 System time out: SIZAR out of band interrupt 841.97 Service image processing unit: output data uncollected 841.98 Service image processing unit: SCNCHN DMA underrun 841.99 Service image processing unit interrupt handler delayed crash 842.00 Scanner failure: communication failure
Remedy: 1 Restart the device, and retry the scan / copy job. Repeat this step with a few copy jobs. Does the error return? Go to step 2. No issue. 2 Is the CCD ribbon cable properly connected to J2 on the controller board? Go to step 3. Properly connect the ribbon cable to J2. 3 Replace the flatbed unit. See “Flatbed removal”. Did this resolve the issue? Problem resolved. Replace the controller board. See “Controller board removal”.