Code: C510
Description: Main charger problem
• MC ALM signal is detected continuously for 800 ms when MC REM signal is turned on.
• MC ALM signal is detected continuously for 800 ms when MC REM signal is turned on.
Causes: Poor contact of the high-voltage transformer PCB connector terminals. Defective high-voltage transformer PCB. Leakage during main charging. Deformed high-voltage transformer PCB terminal spring.
Remedy: Reinsert the connector. Also check for continuity within the connector cable. If none, repair or replace the cable. Replace the high-voltage transformer PCB. Check and clean the main charger assembly. Replace the spring.