Description: RTC error ("Time for maintenance T" appears)
Causes: [Check at start-up]• RTC values are old.• Power has not been turned on for over 5 years.• RTC value is older than 2000/1/1 00:01.[Periodic check per 5 minutes after start-up]• RTC values are older than the ones at the last check.• Partial operation by power reset after C840 error and "Time for Maintenance T" is indicated.
Remedy: 1 Executing U906 The backup battery on the main PWB is faulty, and so, the RTC settings are erased after unplugging the power cord. Execute U906 to reset the indication "Maintenance T". After that, set the data (RTC) via the System Menu.(This treatment needs to be performed when unplugging and reconnecting the power cord.) Date/ Timer/ Energy Saver (System Menu) 2 Replacing the main PWB The main PWB is faulty, or the backup battery runs out. The user call regarding C0840 is frequent even if performing the previous treatment, replace the main PWB. Main PWB Replacement