Description: Error in access to backup data for Reader (reading error at power-on)
Causes: The Reader-related adjustment values could not be read.
Remedy: 1. Clear the backup RAM of RCON. Execute COPIER>FUNCTION>CLEAR>RCON. 2. Enter all the values written on the service label in service mode again. 3. Turn OFF and then ON the main power.