Description: Lamp Error (Scanning) The white level peak did not reach the prescribed threshold when the white plate was scanned.
Causes: • LED defective • IDB (LED driver) defective • SBU defective • IPU defective • Power/signal harness defective • Condensation in scanner unit • Mirrors or lenses dirty or positioned incorrectly • White plate dirty or installed incorrectly
Remedy: 1. Turn the power off/on. 2. Perform the following operations: • Reconnect the power/signal harness. • Reattach/clean the mirrors/lenses. • Reattach/clean the white plate. • Reattach/clean the scanner guide plate. 3. Replace the following parts: • Replace the LED board. • Replace the IDB board or SIO board. • Replace the SBU board. • Replace the IPU board. • Replace the power/signal harness.