Description: IIT Scan Error 62-311-00 IISS LOGIC Fail. Status code 03-557 is raised with this fault code. 62-317-00 IIT Video Over-run 62-318-00 IIT Scanning Timing error 62-319-00 PLL Lock Time Out error 62-321-00 Scan failure 62-322-00 S/W Parameter error 62-360-00 CRG Position Fail 62-345-00 IISS EEPROM Fail 62-362-00 X Hard Fail 62-380-00 AGC Fail 62-386-00 AOC Fail 62-389-00 CRG Over Run Fail NOTE: Status code 14-517 is raised with these fault codes.
Remedy: Cycle system power. The error persists. Y↓N→Go to SCP 6 and complete the Final Actions. Check connections and wiring between the IIT Assembly (Scanner) and IP Board. The wiring is good. Y↓N→Repair or reseat wiring as needed. Install new components as necessary: • IIT Control Board • IIT Assembly • IP Board