Description: MCU Comm Error Communication has failed between the MCU and IP Boards.
Causes: • IP Board, PL7.1.9/7.1.16 • MCU Board, PL7.2.2
Remedy: 1 Reseat the connectors on the IP and MCU Boards then cycle system power. Does the error persist? Go to step 2. Complete. 2 Check the Image Processor Harness continuity. Disconnect the IP Board P/J8 and the MCU Board P/J22. Is the harness damaged? Replace the harness. Go to step 3. 3 Replace the IP Board. Does the error persist? Replace the MCU Board. Complete.