Remedy: Does the fan rotate? • Yes Replace the main board. • No Is the fan connector of the high-voltage power/ sensor board connected? • No Connect the fan connector appropriately. • Yes Does F3-fuse of 86M board open? • Yes Change the F3-fuse. • Yes Replace the fan or main board.
Remedy: Is the cable between 86M board and the high-voltage power supply unit correctly connected? Isn't there defective contact point? Reconnect correctly. Check the defective contact point of high-voltage system. Replace the high voltage power supply.
Remedy: Is the cable between the 86M board and the high-voltage power supply unit correctly connected? Reconnect correctly. Replace the high-voltage power supply.
Description: The dewy of the device is detected Dew condensation occurs easily after the device is carried in from outside. After leaving it at room temperature for 2 hours to half a day, turn on the power again.
Remedy: Dew condensation occurs easily after the device is carried in from outside. After leaving it at room temperature for 2 hours to half a day, turn on the power again. Does the error recur? Yes Turn on the power again after leaving the device for hours. Replace the high-voltage power supply.